Clamp Meter
Required Equipment
7
Required Equipment
Ta b l e 2 is a list of the equipment that is necessary for the performance tests and calibration
adjustment.
Performance Tests
XW
Warning
To prevent possible electrical shock, fire, or personal injury, do not go through the
performance test procedures unless the Product is fully assembled.
The performance tests verify the full operation of the Product and measure the accuracy of
each function against the Product specifications. If the Product fails a part of the test,
calibration adjustment and/or repair is necessary. See Calibration Adjustment.
Before you do the performance tests:
1. Make sure that you have the necessary equipment. See Tabl e 2.
2. Make sure the Product batteries are charged or replace them if necessary.
See Battery Replacement.
3. Warm up the Calibrator as necessary. Refer to its specifications.
4. Let the temperature of the DUT (device under test) become stable to room temperature.
Display and Firmware Version
To verify that all segments of the display function:
1. With the Product off, push and hold H.
2. Turn the Product on.
3. All display segments are shown. See Figure 2.
If segments of the display are missing, repair is necessary. See Contact Fluke.
Table 2. Equipment Requirements
Equipment Required Characteristics Recommended Model
Calibrator 4.5-digit resolution 5522A or equivalent
Wired Coil 50 turns 5500A/Coil or equivalent