Clamp Meter
Specifications
3
Specifications
Accuracy
Accuracy is specified for 1 year after calibration at operating temperatures of 18 °C to 28 °C,
relative humidity a 0 % to 75 %. Accuracy specifications take the form of:
±([% of Reading] + [Number of Least Significant Digits]).
Function Range Resolution
Accuracy
301A 301B 301C 301D 301E
K AC Volts
(45 Hz to 400 Hz)
600.0 V 0.1 V 1.5 %+5
Voltage Frequency
(HZ)
1 Hz to 99.99 kHz
Threshold 10.0 V
9.999 Hz 0.001 Hz
0.1 %+3
99.99 Hz 0.01 Hz
999.9 Hz 0.1 Hz
9.999 kHz 0.001 kHz
99.99 kHz 0.01 kHz
L DC Volts
600.0 V 0.1 V 1 %+5
Ω Resistance
Continuity
Threshold: 30 Ω
600.0 Ω 0.1 Ω
1 %+56.000 kΩ 0.001 kΩ
60.0 kΩ 0.01 kΩ
E Capacitance
9.999 μF 0.001 μF2 %+5
99.99 μF0.01 μF5 %+5
999.9 μF0.1 μF5 %+5
G Diode
3.000 V 0.001 V 10 %
? AC Current A
(45 Hz to 400 Hz)
40.00 A
[2]
0.01 A 2 %+10 NA NA NA NA
400.0 A 0.1 A 2 %+5 NA NA NA NA
60.00 A
[2]
0.01 A NA 2 %+10 2 %+10 2 %+10 2 %+10
600.0 A 0.1 A NA 2 %+5 2 %+5 2 %+5 2 %+5
1000 A1 A NANA2 %+5NA2 %+5
AAC Frequency (Hz)
Threshold 10.00A
45.0 Hz to
400.0 Hz
0.1 Hz 0.1 %+3
A DC Current
[1]
60.00 A
[2]
0.01ANANANA2 % +10
600.00 A 0.1 A NA NA NA 2 % +5
1000 A1 A NANANANA2 % +5
[1] After zero operation.
[2] <1 % range, unspecified.