301A/301B/301C/301D/301E
Calibration Manual
10
Figure 3. AC Current Test Connections
Volts and Ohms
To do the Volts and Ohms performance tests, turn the rotary switch to the necessary function
and apply the values shown in Tab l e 4 . See Figure 4 for the test connections.
0
/
+
5522A
50-Turn Current Coil
Conductor