EasyManua.ls Logo

Fluke 301A - Maintenance and Service

Fluke 301A
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Clamp Meter
Performance Tests
9
Table 3. Current Performance Tests
Value Unit
Frequency
or
Amplitude
Res. Spec. Counts
301 Model
Lower
Limit
Upper
Limit
ABCDE
A ac (with 50-turn coil)
0.2
A50 Hz
0.01
2 %
10
•••••
9.7 10.3
0.8 0.01 10
•••••
39.10 40.90
20.15
•••••
97.5 102.5
80.15
391.5 408.5
12 0.1 5
••••
587.5 612.5
14 1
5
••
681 719
19.5 1 5
••
951 1000
A ac, Hz (with 50-turn coil)
45.0
Hz
0.2 A
0.1 0.1 %
3
•••••
44.7 45.3
400.0 3
•••••
399.3 400.7
45.0
8 A
3
•••••
44.7 45.3
400.0 3
•••••
399.3 400.7
A dc (with 50-turn coil) must zero for each point
-16
A0 Hz
1
2 %
5
-821 -779
-6 0.1 5
••
-306.5 -293.5
-0.6 0.01 10
••
-30.70 -29.30
0.2 0.01 10
••
9.70 10.30
1.2 0.01 10
••
58.7 61.3
20.1
5
••
97.5 102.5
12 0.1 5
••
587.5 612.5
14 1 5
681 719
19.5 1 5
951 1000

Other manuals for Fluke 301A