302+/303/305
Calibration Manual
8
Jaw Maintenance
If the Product does not work properly:
1. Inspect the jaw mating surface for cleanliness. If any foreign material (including rust)
is present, the jaw will not close properly and measurement errors will result.
2. Open the jaws and wipe the clamp metal ends with a non-flammable oil and cloth.
User-Replaceable Parts
User-replaceable parts are shown in Table 2. To order parts, see “Contact Fluke”.
Table 2. User-Replaceable Parts
Fluke Part
Number
Description Quantity
3986568 FLUKE-CAP-2001,TL7X PROBE CAP,BLACK 1
3986579 FLUKE-CAP-2001-01,TL7X PROBE CAP,FLUKE RED 1
2444986 BATTERY 1.5V MICRO ALKALINE IEC LR03 AAA 2
4087581 BATTERY DOOR (China only) 1
4198591 BATTERY DOOR (Brazil only) 1
3765741 FOAM BAG 1
4045130 302+/303 USERS MANUAL 1
4045148 305 USERS MANUAL 1
Required Equipment
The equipment in Table 3 is necessary for the performance tests and calibration
adjustment.
Table 3. Required Equipment
Equipment Required Characteristics Recommended Model
Calibrator 4.5-digit resolution FLUKE 5522A or equivalent
Wired coil 50 turns
FLUKE 5500A/Coil or
equivalent