Clamp Meter
Performance Tests
11
Temperature, Volts, Ohms, Frequency, and Capacitance
To do the temperature, volts, ohms, and capacitance performance tests:
1. Connect the Calibrator NORMAL output and ground to the Product. See Figure 4 for
the connections.
2. Apply the values shown in Table 4.
If the Product shows indications outside of the limits shown in Table 4, calibration
adjustment is necessary. See “Calibration Adjustment”.
0
/
+
300V
CAT
600VCAT
325
TRUE RMS CLAMP METER
5522A
hfk07.eps
Figure 4. Temperature, Volts, Ohms, Frequency, and Capacitance Performance Test Connections