323/324/325
Calibration Manual
12
Current
To do the current tests:
1. Connect the Calibrator current output and ground to the 50-Turn Coil. See Figure 5.
2. Turn the Product to the necessary current mode.
3. Apply the input level for each step shown in Table 4.
4. Compare the indication on the Product display with the UUT reading limits in
Table 4.
5. If the Product shows indications outside of the limits shown in Table 4, calibration
adjustment is necessary. See “Calibration Adjustment”.
0
•
/
+
5522A
50-Turn Coil
Conductor
hfk025.eps
Figure 5. AC Current Test Connections