EasyManua.ls Logo

Fluke 377 - Page 20

Fluke 377
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
377/377 FC/378/378 FC
Users Manual
16
3. Keep the probe coupling >2.5 cm (1 inch) away from the
conductor.
4. Turn the control knob to
K.
The display shows
J to indicate that the measurements are
from the iFlex Probe. When the current measurement is
<0.5 A, the center dot in the icon flashes. For current
measurements >0.5 A, the center dot is steady.
The display shows the measurement.
If the iFlex Probe does not work as expected:
Make sure that the coupling system is connected and closed
correctly or look for any damage. If any foreign material is
present, the coupling system will not close properly.
Inspect the cable between the iFlex Probe and the Clamp for
any damage.
Check that the control knob is in the correct position (
K).
Figure 3. iFlex Probe Setup
<0.5 A
>0.5 A
1 2 3 4
>2.5 cm
>1 in

Other manuals for Fluke 377

Related product manuals