3 Fluke Corporation Solar Tools Kit with 393 FC Clamp Meter, Irradiance Meter and Solar Test Leads
https://www.fluke.com/en-us/product/electrical-testing/clamp-meters/solar-tools-393-irrsol
Distance from Optimum i2500-10 Flex i2500-18 Flex Error
A 0.5 in (12.7 mm) 1.4 in (35.6 mm) ±0.5 %
B 0.8 in (20.3 mm) 2.0 in (50.8 mm) ±1.0 %
C 1.4 in (35.6 mm) 2.5 in (63.5 mm) ±2.0 %
Measurement uncertainty assumes centralized primary conductor at optimum position, no external electrical or magnetic field, and within operating temperature
range.
DC Current
Range 999.9 A
Resolution 0.1 A
Accuracy
2 % RD + 5 digits
[1]
[1]
When using the ZERO (B) function to compensate for offsets.
AC Voltage
Range 600.0 V
1000 V
Resolution 0.1 V (≤600.0 V)
1 V (≤1000 V)
Accuracy 1 % RD + 5 digits (20 Hz to 500 Hz)
DC Voltage
Range 600.0 V
1500 V
Resolution 0.1 V (≤600.0 V)
1 V (≤1500 V)
Accuracy 1 % RD + 5 digits
mV dc
Range 500.0 mV
Resolution 0.1 mV
Accuracy 1 % RD + 5 digits
Amps Frequency: Jaw
Range 5.0 Hz to 500.0 Hz
Resolution 0.1 Hz
Accuracy 0.5 % RD + 5 digits
Trigger Level 5 Hz to 10 Hz, ≥10 A
10 Hz to 100 Hz, ≥5 A
100 Hz to 500 Hz, ≥10 A
Amps Frequency: Flexible Current Probe
Range 5.0 Hz to 500.0 Hz
Resolution 0.1 Hz
Accuracy 0.5 % RD + 5 digits
Trigger Level 5 Hz to 20 Hz, ≥25 A
20 Hz to 100 Hz, ≥20 A
100 Hz to 500 Hz, ≥25 A