6
Functions
Fluke 416D
GB
D
F
I
E
P
NL
DK
NO
S
FIN
GR
TR
CZ
SK
PL
HU
RU
RO
CN
KOR
Functions
Addition / Subtraction
Distance Measurement.
1x briefly: the next measurement is added to
the previous one.
2x briefly: the next measurement is subtracted
from the previous one.
Repeat this procedure for as many times as required.
The result is displayed in the primary display, the
previously measured value is displayed in secondary
display line 2, the value to be added in secondary
display line 1.
1x briefly: the last step is canceled.
Area function
1x briefly: the symbol is displayed.
1x briefly: takes first distance measurement
(e.g. length)
1x briefly: takes second distance measurement
(e.g. width)
The result of the area measurement is displayed in
the primary display, the individually measured values
are displayed in secondary display lines 1 and 2.
Adding and subtracting areas
Calling up the area function and measuring areas.
Press once/twice.
1x briefly: takes first distance measurement
(e.g. length)
1x briefly: takes second distance measurement
(e.g. width)
The result of the second area measurement, "+"
flashes.
1x briefly: confirms the addition; the added area
results are displayed in the primary display.
Volume function
2x briefly: the symbol is displayed.
1x briefly: takes first distance measurement
(e.g. length)
1x briefly: takes second distance measurement
(e.g. width)
The result of the area measurement from the values
already measured is displayed in the primary display.
1x briefly: takes the third distance measure-
ment (e.g. height). The value is displayed in
secondary display line 1.
The result of the area measurement is displayed in
the primary display, the two previously measured

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 416D and is the answer not in the manual?

Fluke 416D Specifications

General IconGeneral
BrandFluke
Model416D
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals