5730A
Getting Started Manual
36
DC Current Secondary Performance Specifications and Operating Characteristics
Range
Stability
±1 °C
[1]
24 Hours
Temperature
Coefficient
[2]
Compliance
Limits
Burden
Voltage
Adder
[3]
(±nA/V)
Maximum
Load for
Published
Specification
[4]
(Ω)
Noise
10 - 40 °C
0 - 10 °C
and
40 - 50 °C
Bandwidth
0.1-10 Hz
Bandwidth
10 Hz-10 kHz
pk-pk RMS
±(ppm
output + nA)
±(ppm output + nA) /
°C
ppm output
+ nA
nA
220 μA
2.2 mA
22 mA
220 mA
2.2 A
5 + 1
5 + 5
5 + 50
8 + 300
9 + 7 μA
1 + 0.40
1 + 2
1 + 20
1 + 200
1 + 2.5 μA
3 + 1
3 + 10
3 + 100
3 + 1 μA
3 + 10 μA
10
10
10
10
3
[5]
0.2
0.2
10
100
2 μA
20k
2k
200
20
2
6 + .9
6 + 5
6 + 50
9 + 300
12 + 1.5 μA
10
10
50
500
20 μA
5725A
±(ppm
output + μA)
±(ppm output + μA)/
°C
ppm output
+ μA
μA
11 A 25 + 100 20 + 75 30 + 120 4 0 4 15 + 70 175
Notes:
Maximum output from the calibrator’s terminals is 2.2 A. Specifications for 220 μA and 2.2 mA ranges are increased by a factor of 1.3
when supplied through 5725A terminals.
1. Stability specifications are included in the Absolute specification values for the primary specifications.
2. Temperature coefficient is an adder to Absolute specifications. It does not apply unless operating more than ±5 °C from
calibration temperature.
3. Burden voltage adder is an adder to Absolute specifications that does not apply unless burden voltage is greater than 0.5 V.
4. For higher loads, multiply Absolute specification by:
ionspecificat publishedfor load maximum
load actualx 0.1
1 +
5. The calibrator’s compliance limit is 2 V for outputs from 1 A to 2.2 A. 5725A Amplifier may be used in range-lock mode down to
0 A.
Minimum Output .................................................. 0 for all ranges, including 5725A.
Settling Time ........................................................ 1 second for μA and mA ranges; 3 seconds for 2.2 A range; 6 seconds
for 11 range; + 1 second for range or polarity change
Overshoot ............................................................. <5 %

Table of Contents