Multifunction Calibrator
Electrical Specifications
37
AC Current Specifications
5730A AC Current Specifications: 99 % Confidence Level
Range Resolution
Frequency
(Hz)
Absolute / ±5 °C from calibration temperature Relative ±1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
±(ppm output + nA)
220 μA
1 nA
10 - 20
20 - 40
40 - 1 k
1 k - 5 k
5 k - 10 k
260 + 20
170 + 12
115 + 10
300 + 15
1000 + 80
280 + 20
180 + 12
117 + 10
320 + 15
1100 + 80
290 + 20
190 + 12
118 + 10
340 + 15
1200 + 80
300 + 20
200 + 12
120 + 10
350 + 15
1300 + 80
260 + 20
130 + 12
100 + 10
250 + 15
900 + 80
280 + 20
150 + 12
110 + 10
280 + 15
1000 + 80
2.2 mA 10 nA
10 - 20
20 - 40
40 - 1 k
1 k - 5 k
5 k - 10 k
260 + 50
170 + 40
115 + 40
210 + 130
1000 + 800
280 + 50
180 + 40
117 + 40
220 + 130
1100 + 800
290 + 50
190 + 40
118 + 40
230 + 130
1200 + 800
300 + 50
200 + 40
120 + 40
240 + 130
1300 + 800
260 + 50
130 + 40
100 + 40
190 + 130
900 + 800
280 + 50
150 + 40
110 + 40
220 + 130
1000 + 800
22 mA 100 nA
10 - 20
20 - 40
40 - 1 k
1 k - 5 k
5 k - 10 k
260 + 500
170 + 400
115 + 400
210 + 700
1000 + 6000
280 + 500
180 + 400
117 + 400
220 + 700
1100 + 6000
290 + 500
190 + 400
118 + 400
230 + 700
1200 + 6000
300 + 500
200 + 400
120 + 400
240 + 700
1300 + 6000
260 + 500
130 + 400
100 + 400
190 + 700
900 + 6000
280 + 500
150 + 400
110 + 400
220 + 700
1000 +
6000
±(ppm output + μA)
220 mA
1 μA
10 - 20
20 - 40
40 - 1 k
1 k - 5 k
5 k - 10 k
260 + 5
170 + 4
115 + 3
210 + 4
1000 + 12
280 + 5
180 + 4
117 + 3
220 + 4
1100 + 12
290 + 5
190 + 4
118 + 3
230 + 4
1200 + 12
300 + 5
200 + 4
120 + 3
240 + 4
1300 + 12
260 + 5
130 + 4
100 + 3
190 + 4
900 + 12
280 + 5
150 + 4
110 + 3
220 + 4
1000 + 12
2.2 A
10 μA
20 - 1 k
1 k - 5 k
5 k - 10 k
270 + 40
440 + 100
6000 + 200
280 + 40
460 + 100
7000 + 200
290 + 40
480 + 100
7500 + 200
300 + 40
500 + 100
8000 + 200
260 + 40
420 + 100
6000 + 200
280 + 40
440 + 100
7000 + 200
5725A Amplifier:
11 A
100 μA
40 - 1 k
1 k - 5 k
5 k - 10 k
370 + 170
800 + 380
3000 + 750
400 + 170
850 + 380
3300 + 750
440 + 170
900 + 380
3500 + 750
460 + 170
950 + 380
3600 + 750
300 + 170
700 + 380
2800 + 750
330 + 170
800 + 380
3200 + 750
Note:
Maximum output from the calibrator’s terminals is 2.2 A. Specifications for 220 μA and 2.2 mA ranges are increased by a factor of 1.3
plus 2 μA when supplied through 5725A terminals.

Table of Contents

Other manuals for Fluke 5730A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5730A and is the answer not in the manual?

Fluke 5730A Specifications

General IconGeneral
TypeMultifunction Calibrator
Resistance Range0 to 1 GΩ
AC Voltage Frequency10 Hz to 1 MHz
DisplayColor LCD
InterfacesRS-232, Ethernet
Weight18 kg
DC Voltage Range0 to ±1100 V
Dimensions172 mm H x 432 mm W x 457 mm D (6.8 in H x 17 in W x 18 in D)