Multifunction Calibrator
Electrical Specifications
45
Note
The specifications for these coils are at 99 % confidence level and are the combined
specification of the coil and a 52120A. If the coils are used with other current
sources the calibration specification of the coils alone is 0.65 % (99 % confidence
level) from 0 Hz to 10 kHz.
Operating Limits
Output Current Range
2 A 20 A 120 A
Current Output (Max.) 2 A rms 20 A rms 120 A rms
Current Input
Input Current (Max.) 200 mA rms 200 mA rms 120 mA rms
Current gain 10 100 1,000
Voltage Input
Input Voltage (Max.) 2 V rms 2 V rms 1.2 V rms
Transconductance 1 Siemen 10 Siemens 100 Siemens
120 A Range Current/Frequency Limits
Frequency Maximum Output Current Maximum Current Input Maximum Voltage Input
DC
±100 A ±100 mA ±1.0 V
<10 Hz 100 A pk (70 A rms) 100 mA pk (70 mA rms) 1.0 V pk (0.7 V rms)
10 Hz to 10 kHz 170 A pk (120 A rms) 170 mA pk (120 mA rms) 1.7 V pk (1.2 V rms)
Note:
The 2 A and 20 A ranges operate at full output current from DC to 10 kHz.
Output Isolation
Frequency Maximum Voltage Signal Applied to any Output Current Terminal with respect to Earth
DC to 850 Hz 600 V rms, 850 V pk, limited 2 A rms, no transient overvoltages
850 Hz to 3 kHz 100 V rms, 142 V pk, limited 2 A rms, no transient overvoltages
3 kHz to 10 kHz 33 V rms, 47 V pk, limited 2 A rms, no transient overvoltages