Electrical Power Standards
6100B Set Up Exercise
17
Sample Reference Signal and F10
The external keyboard F10 key replicates the “OUTPut:ROSCillator” ON/OFF GPIB
command. Pushing F10 toggles the Sample Reference signal output On and Off.
The Sample Reference signal can be used to synchronize sampling instrumentation in a
measurement system. This is described in the Users Manual.
6100B Set Up Exercise
A 6100B or 6105A is required to continue with this training exercise. The task is to set up
the current channel to output 2.0 A. After following the Power Up instructions earlier in
this manual, the L1 voltage channel (L1 V) output values are shown with a white
background. See Figure 4. Note the L2, L3 and N channels are grayed out as Auxiliary
units are not connected.
gdw08.bmp
Figure 4. The Output Menu (L1 Voltage Selected)
This exercise is divided into seven sections:
1. Switch to the Current Output
2. Enable the Current Channel
3. The Current Channel is Enabled but not Operating
4. Set up the L1 Current Channel
5. Over Compliance Warning Message
6. Complete the Current Circuit and Turn the Output On
7. Turn the Output Off
An explanation of each of these follows.