Fluke 705
Calibration Information
2
Specifications
Product specifications are provided Tables 1 and 2. These specifications are based on a 1-year calibration cycle and
apply from 18 ºC to 28 ºC (64 ºF to 82 ºF) unless stated otherwise. Accuracy specifications are given as follows:
± ( [ % of reading ] + [ number of least significant digits ] ).
Table 1. General Specifications
Maximum voltage between any
terminal and earth ground or
between any two terminals
30 V
Display: 0 to 28 V with 0.001 V resolution, 0 to 24 mA with 0.001 mA resolution
Operating Temperature: -10 °C to 55 °C
Storage Temperature: -40 °C to 60 °C
Temperature Coefficient: ± 0.005 % of range per °C for -0 °C to 18 °C and 28 °C to 55 °C
Relative Humidity: 95 % to 30 °C,
75 % to 40 °C,
45 % up to 50 °C,
35 % up to 55 °C
Operationing Altitude: 3000 meters maximum
Power Requirements: Single 9 V battery (ANSI/NEDA 1604A or IEC 6LR61)
Battery Life (Typical): Source Mode: 18 hours; 12 mA into 500 Ω
Measure / Simulate Mode: 50 hours
Shock, Vibration: 1 meter drop test
Random vibration 2 g @ 5 to 500 Hz
Size (H x W x L): 32 mm x 87 mm x 187 mm (1.25 in x 3.41 in x 7.35 in)
Weight: 349 g (12.3 oz). With Holster and flex-stand: 601 g (21.2 oz)
Safety Certified as compliant to CAN/CSA C22.2 No. 1010.1:1992.
Complies with ANSI/ISA S82.01-1994
Certification P and
)
Table 2. Accurcy Specifications
Function Range Resolution Accuracy
DC Volts Input + 28 V 0.001 V ± (0.025 % + 1)
DC mA Input 24 mA 0.001 mA ±( 0.02 % + 2)
DC mA Output 0 mA to 24 mA 0.001 mA ± (0.02 % + 2)
E 0 % -100 % - 0 % 40 second smooth ramp
O 0 % -100 % - 0 % 15 second smooth ramp
N 0 % -100 % - 0 % Stair-step ramp in 25 % steps, pausing 5 seconds at each step
Source mode: Compliance: ≥ 1000 Ω at 20 mA
Simulate mode: External loop voltage requirements: 24 V nominal, 30 V maximum, 12 V minimum
Loop Power ≥ 24 V
Percent display - 25 % to 125 %
Input/Output protection Input/Output protection on mA range: resettable, nonreplaceable 0.1 A fuse.