EasyManua.ls Logo

Fluke 707 - mA Sourcing Modes; Using mA Sourcing

Fluke 707
8 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibrating voltage input
signal conditioners
Fluke loop calibrators are ideal
for calibrating many 4-20 mA
signal conditioners using their
precision current sourcing and
simulation capabilities. However,
there are many signal condi-
tioners that require a precision
voltage source for proper cali-
bration. Using a simple
precision resistor and standard
connector, Fluke loop calibra-
tors can field calibrate many
standard and non-standard
voltage input signal condition-
ers. This approach works well
for the 705, 707 and 787 loop
calibration tools (this is not
necessary with the 715 loop
calibrator as it has a precision
direct voltage output).
Voltage input signal condi-
tioners come in many varieties.
The most common are 0-10 V,
0-5 V and 1-5 V input levels.
Typical outputs of these devices
are an isolated or non-isolated
0-10 V or 4-20 mA. Resistors
with values of 250 to 500
are common loop load resistors
and provide voltage input lev-
els as a function of the loop
current (see Figure 4).
Using Fluke loop calibrators
as a voltage source
A precision shunt resistor may
be used to derive voltages for
calibration using the calibra-
tor’s current source mode.
Using this system, Fluke loop
calibrators are capable of gen-
erating voltages for devices
with input spans as low as 10
mV to as high as 24 V. Table 1
gives values of precision resis-
tors to accommodate a variety
of voltage calibrations and the
Fluke calibrator ideal for each
application.
Constructing a precision
load resistor assembly
A simple precision current
shunt can be constructed using
a precision RN60, 1 W resistor,
(see table for the correct value),
a dual banana jack connector
and some test leads with alli-
gator clips (see Figure 5). The
RN60 class resistor is available
from many commercial sources.
Construct the assembly as
shown in Figure 5. This preci-
sion resistor assembly, coupled
with the precision current
sourcing capabilities of Fluke
loop calibrators, generates pre-
cision voltages to cover 1-5 V
or 2-10 V applications. The
1,000 assembly has an
advantage in that it allows a
direct one-to-one display cor-
relation to voltage when
sourcing current from the Fluke
loop calibrator during calibra-
tion (1 mA = 1 V). The following
example will utilize a 250 to
take advantage of the 0-100%
and 25% buttons on the Fluke
707 (25% or 4 mA = 1 V).
4 Fluke Corporation Troubleshooting and Maintenance Using Fluke Loop Calibrators
Figure 4
Resistor Sourced Generated Fluke Loop
Value Application Current Value Voltage Calibrator
10 mV 0-20.000 mA 0-200.00 mV 707, 705, 787
transducer
250 0-5 V Source 0-20.000 mA 0-5.000 V 705, 707
1000 0-10 V Source 0-12.000 mA 0-12.000 V 705, 707, 787
Note: Input impedance of the device under test should be 1 M
Table 1. Values of precision resistors to accommodate a variety of voltage calibrations
Figure 5
Constructing a precision current shunt
Parts required:
1. Resistor: 1000
Type: RN60 (50 ppm@70 °C) 1 W
2. 36˝ test leads with alligator clips
3. Connector: Dual Banana Plug 2/screw terminals

Other manuals for Fluke 707

Related product manuals