73 Series III Instruction Sheet Supplement
8/04 1
Change #1
On the first panel, under Warning, add the following:
Do not use this meter in a manner not
specified by the instruction sheet or the
protection provided by the meter may be
impaired.
Under General Specifications, change the following:
From: Shock, Vibration per MIL-T-PRF 28800F
Class III, Sinusoidal,
Non Operating.
To: Vibration per MIL-PRF-28800 for
a Class II instrument.
Shock 1 Meter drop per IEC
61010-1:2001.
From: Safety 600 VCAT II per
ANS/ISA S82.01-1994,
EN 61010-1: 1993, CSA
C22.2 No 1010.1-92,
UL 3111-1.
To: Safety Complies with ANSI/ISA
S82.01-2004, CSA 22.2
No. 1010.1:2004 to 600
V Overvoltage Category
II, listed to UL61010-1.
Licensed by TÜV to
EN61010-1.
Change #2
Under Continuity Test ( R ):
Change: (resistance <210 e)
To: (resistance <150 e)