Do you have a question about the Fluke 700G Series and is the answer not in the manual?
Ingress Protection | IP40 |
---|---|
Accuracy | 0.05% of full scale |
Battery Life | 1500 hours typical |
Operating Temperature | -10°C to +55°C |
Communication | USB |
Weight | Approximately 1.5 lbs (0.68 kg) |
Pressure Media | Non-corrosive gases and liquids |
Pressure Connection | 1/4 in NPT female |
Storage Temperature | -40°F to 158°F (-40°C to 70°C) |
Pressure Units | psi, bar, kPa |
Resolution | Varies by model |
Warranty | Three years |
Power | 3 AA batteries |
Provides telephone numbers and website for contacting Fluke for support.
Details safety requirements and approvals for hazardous locations and electrical hazards.
Outlines specific usage conditions and warnings to prevent potential hazards.
Explains the meaning of various symbols used on the product and in the manual.
Lists the pressure ranges available for the different models of the gauge.
Details the accuracy specifications for the 700G series pressure gauges.
Details the accuracy specifications for the 700RG series pressure gauges.
Specifies the types of gases and liquids compatible with the product.
Covers operating and storage temperature, humidity, and IP rating.
Addresses compliance with electromagnetic wave equipment standards for industrial use.
Provides dimensions, connection type, housing material, display, and power details.
Presents a detailed table of pressure ranges and resolutions across various engineering units for different models.
Instructions for cleaning the pressure gauge using a soft cloth and mild soap.
Step-by-step guide on replacing the AA batteries, including safety warnings.
Lists the necessary equipment for performing product verification tests.
Detailed procedure for verifying pressure measurements using test points and equipment.
Table listing specific pressure points and their acceptable tolerance limits for verification.
Specifies the required pressure/vacuum standards and TUR for calibration.
Details the NPT male connection and necessary adapters for pressure standards.
Instructions for setting up serial interface communication for calibration adjustments.
Describes the multi-point calibration procedure for non-linear error correction.
Illustrates the calibration steps using a specific model (700G07) as an example.
Table detailing specific calibration points for various 700G/GA/RG models.
Explains terminal communication setup and parameters for connecting a PC to the product.
Table detailing measurement units that can be displayed or set via serial port commands.