700G Series
Calibration Manual
20
Table 4. Calibration Points (Cont.)
Model
Number
700G10 700G29 700G30 700G31 700GA4 700GA5 700GA6 700GA27
Calibration
Point 2 1750.0 2625.0 4375.0 9300 13.125 26.250 87.50 262.50
Calibration
Point 3 1500.0 2250.0 3750.0 8000 11.000 22.500 75.00 225.00
Calibration
Point 4 1250.0 1875.0 3125.0 6500 25.000 18.750 62.50 187.50
Calibration
Point 5 1000.0 1500.0 2500.0 5000 9.375 15.000 50.00 150.00
Calibration
Point 6 750.0 1125.0 1875.0 3900 7.500 11.250 37.50 112.50
Calibration
Point 7 500.0 750.0 1250.0 2800 5.625 7.500 25.00 75.00
Calibration
Point 8 250.0 375.0 625.0 1800 3.750 3.750 12.50 37.50
Calibration
Point 9 0.0 0.0 0.0 1000 1.875 0.300 0.40 0.50
Calibration
Point 10 500 0.300
Calibration
Point 11 0
Serial Interface
Use terminal communication software on a PC to set up terminal communication. An
RS-232 to USB cable is necessary. This cable comes with 700G/TRACK Software.
Use these terminal parameters:
• Bits per second: 9600
• Data bits: 8
• Parity: None
• Stop bits: 1
• Flow control: None
• Local echo: on