Pressure Gauge
Calibration Adjustment
19
Table 4. Calibration Points
Model
Number
700G01 700G02 700G04 700G05 700G05 700G27 700G07 700G08
Pressure
Range (psi) 0.4 1 15 30 100 300 500 1000
Calibration
Point 1 0.3600 1.0000 15.000 30.000 100.00 300.00 500.00 1000.0
Calibration
Point 2 0.2880 0.8000 13.000 25.500 87.50 262.50 437.50 875.0
Calibration
Point 3 0.2080 0.6000 10.750 21.500 75.00 225.00 375.00 750.0
Calibration
Point 4 0.1280 0.4000 8.500 17.000 62.50 187.50 312.50 625.0
Calibration
Point 5 0.0480 0.2000 6.500 12.750 50.00 150.00 250.00 500.0
Calibration
Point 6 0.0000 0.0000 4.300 8.500 37.50 112.50 187.50 375.0
Calibration
Point 7 -0.0900 -0.2500 2.150 4.250 25.00 75.00 125.00 250.0
Calibration
Point 8 -0.1800 -0.5000 0.000 0.000 12.50 37.50 62.50 125.0
Calibration
Point 9 -0.2700 -0.7500 -7.000 -7.000 0.00 0.00 0.00 0.0
Calibration
Point 10 -0.3600 -1.0000 -14.000 -14.000 -12.00 -12.00 -12.00 -12.0
Calibration
Point 11
Model
Number
700G10 700G29 700G30 700G31 700GA4 700GA5 700GA6 700GA27
Pressure
Range (psi) 2000 3000 5000 10000 15 psia 30 psia 100 psia 300 psia
Calibration
Point 1 2000.0 3000.0 5000.0 10000 15.000 30.000 100.00 300.00