700G Series
Calibration Manual
18
push to continue. As the Product measures, the screen will show ———-. When
measurements are complete the screen will show the calibration value.
4. Use the Pressure Standard to output 312.50 psi (62.5 %). After the output is stable,
push to continue. As the Product measures, the screen will show ———-. When
measurements are complete the screen will show the calibration value.
5. Use the Pressure Standard to output 250.00 psi (50 %). After the output is stable,
push to continue. As the Product measures, the screen will show ———-. When
measurements are complete the screen will show the calibration value.
6. Use the Pressure Standard to output 187.50 psi (37.5 %). After the output is stable,
push to continue. As the Product measures, the screen will show ———-. When
measurements are complete the screen will show the calibration value.
7. Use the Pressure Standard to output 125.00 psi (25 %). After the output is stable,
push to continue. As the Product measures, the screen will show ———-. When
measurements are complete the screen will show the calibration value.
8. Use the Pressure Standard to output 62.50 psi (12.5 %). After the output is stable,
push to continue. As the Product measures, the screen will show ———-. When
measurements are complete the screen will show the calibration value.
9. Use the Pressure Standard to output 0.00 psi. After the output is stable, push to
continue. As the Product measures, the screen will show ———-. When
measurements are complete the screen will show the calibration value.
Note
Only some ranges use vacuum calibration. If your Product does not,
then this step can be skipped and calibration is complete.
10. Use the Pressure Standard to output -12.00 psi. After the output is stable, push to
continue. As the Product measures, the screen will show ———-. When the
measurements are complete, a Product reset occurs and the Product turns off then
turns on as usual.