Fluke 753 Specifications

General IconGeneral
TypeDocumenting Process Calibrator
Measurement FunctionsVoltage, Current, Resistance, Frequency, Temperature, Pressure
Source FunctionsVoltage, Current, Resistance, Frequency, Temperature, Pressure
Current DC Range0 to 24 mA
Current DC Resolution0.001 mA
Current AC Resolution0.001 mA
Resistance Resolution0.1 Ω
Frequency Resolution0.001 Hz
Temperature ThermocoupleJ, K, T, E, R, S, B, L, U, N, C
HART CommunicationYes
Loop Power24 V
DisplayGraphic LCD
InterfaceRS-232
Power SupplyRechargeable battery
Voltage DC Range0 to 30 V
Voltage DC Accuracy±(0.02% of reading + 2 counts)
Voltage AC Range0 to 20 V
Voltage AC Resolution0.01 mV
Current DC Accuracy±(0.02% of reading + 2 counts)
Resistance Range0 to 3200 Ω
Frequency Accuracy±0.01% of reading
Temperature RTDPt100, Pt200, Pt500, Pt1000
Operating Temperature-10 °C to 50 °C
Storage Temperature-20 °C to 60 °C