754
Users Guide
6
Supported vs. Generic Transmitters
The Product communicates with most HART transmitters. “Supported Transmitters” are transmitters in which the Product is
programmed to use their device-dependent commands and software version. “Generic transmitters” are transmitters that are
not programmed in the Product. Table 1 shows the operations that are available for supported vs. generic transmitters in
single-point and multidrop configurations.
Table 1. Supported vs. Generic Transmitters (Devices)
Supported Transmitter Generic Transmitter
Menu Operation
Single Point Multidrop Single Point Multidrop
Top Level Active Device screen
Basic (read/write, cloning capability)
Sensor
Read only Read only
(Temperature) Sensor (read/write)
Device Identification (read/write)
HART Output (read/write)
Setup
HART Information (read only)
Loop Test Not Available Not Available
Pressure Zero Trim
Output Trim Not Available Not Available
Service
Sensor Trim
Not Available Not Available
Process Detailed process information