Documenting Process Calibrator
Detailed Specifications
5
DC Voltage Measurement
Range Resolution
% of Reading + Floor
1-Year 2 Year
±3.00000 V
0.00001 V 0.02 % + 0.00005 V 0.03 % + 0.00005 V
±30.0000 V
0.0001 V 0.02 % + 0.0005 V 0.03 % + 0.0005 V
±300.00 V
0.01 V 0.05 % + 0.05 V 0.07 % + 0.05 V
Input Impedance: >4 MΩ
Maximum Input Voltage: 300 V, IEC 61010 300V CAT II
Temperature coefficient: (0.001 % of reading + 0.0002 % of range) / °C (<18 °C or >28 °C)
Normal mode rejection: >100 dB at 50 or 60 Hz nominal
AC Voltage Measurement
Range
40 Hz – 500 Hz
Resolution
% of Reading + Floor
1-Year 2 Year
3.000 V 0.001 V 0.5 % + 0.002 V 1.0 % + 0.004 V
30.00 V 0.01 V 0.5 % + 0.02 V 1.0 % + 0.04 V
300.0 V 0.1 V 0.5 % + 0.2 V 1.0 % + 0.2 V
Input Impedance: >4 M Ω and <100 pF
Input Coupling: AC
Maximum Input Voltage: 300 V, IEC 61010 300V CAT II
Temperature coefficient: 5 % of specified accuracy / °C (<18 °C or >28 °C)
Specifications apply for 9 % to 100 % of voltage range.
DC Current Measurement
Range Resolution
% of Reading + Floor
1-Year 2 Year
±30.000 mA 1 μA
0.01 % + 5 μA 0.015 % + 7 μA
±100.00 mA 10 μA
0.01 % + 20 μA 0.015 % + 30 μA
Maximum Input: 110 mA
Maximum Burden Voltage: 420 mV at 22 mA
Temperature coefficient: 3 % of specified accuracy / °C (<18 °C or >28 °C)
No Fuse
Normal mode rejection: 90 dB at 50 or 60 Hz nominal, and 60 dB at 1200 Hz and 2200 Hz nominal (HART signals)
Resistance Measurement
Range Resolution
% of Reading + Floor
Source
Current
1-Year 2 Year
10.000 Ω 0.001 Ω 0.05 % + 0.050 Ω 0.07 % + 0.070 Ω
3 mA
100.00 Ω 0.01 Ω 0.05 % + 0.05 Ω 0.07 % + 0.07 Ω
1 mA
1.0000 kΩ 0.1 Ω 0.05 % + 0.0005 kΩ 0.07 % + 0.0007 kΩ 500 μA
10.000 kΩ 1 Ω 0.10 % + 0.010 kΩ 0.15 % + 0.015 kΩ 50 μA
Open-circuit voltage: 5 V nominal
Temperature coefficient: 3 % of specified accuracy / °C (<18 °C or >28 °C)
Continuity Testing
Tone Resistance
Continuous tone
May or may not get tone
No tone
<25 Ω
25 to 400 Ω
>400 Ω