753/754
Calibration Manual
12
AC Volts Measurement
To verify the ac volts measurement function:
1. Connect the UUT to the 5522A as shown in Figure 1.
2. Set the UUT to the ac volts measurement function.
3. Push the Range softkey on the UUT to lock on the 3.0 V range.
4. Set the 5522A to the first test point in Table 4.
5. Stop to let the output become stable.
6. See if the value shown on the UUT is in the range shown in the applicable
column.
7. Continue through the test points. Make sure to lock the UUT on the specified
range.
8. When you complete the test, set the 5522A to STANDBY.
Table 4. AC Volts Measurement Verification Points
UUT
Ran
e
Input
RMS
Frequency
Minimum
1-Yea
Maximum
1-Yea
Minimum
2-Yea
Maximum
2-Yea
3.000 V 0.26 500 Hz 0.257 0.263 0.253 0.267
3.000 V 3 500 Hz 2.983 3.017 2.966 3.034
3.000 V 0.26 40 Hz 0.257 0.263 0.253 0.267
3.000 V 3 40 Hz 2.983 3.017 2.966 3.034
30.00 V 2.6 500 Hz 2.567 2.633 2.53 2.67
30.00 V 30 500 Hz 29.830 30.170 29.66 30.34
30.00 V 2.6 40 Hz 2.567 2.633 2.53 2.67
30.00 V 30 40 Hz 29.830 30.170 29.66 30.34
300.0 V 27 500 Hz 26.665 27.335 26.5 27.5
300.0 V 295 500 Hz 293.325 296.675 291.9 298.2
300.0 V 27 40 Hz 26.665 27.335 26.5 27.5
300.0 V 295 50 Hz 293.325 296.675 291.9 298.2