Documenting Process Calibrator
Performance Verification Tests
15
Fluke 75X
Fluke 5522A
2-WIRE COMP ON
gso05.eps
Figure 3. Resistance Measurement Verification Connections
Table 6. Resistance Measurement Verification Points
UUT
Range
Input
Minimum
1-Year
Maximum
1-Year
Minimum
2-Year
Maximum
2-Year
10.000 Ω 0 Ω -0.050 0.050 -0.070 0.070
10.000 Ω 10 Ω 9.945 10.055 9.923 10.077
100.00 Ω 0 Ω -0.05 0.05 -0.07 0.07
100.00 Ω 100 Ω 99.90 100.10 99.86 100.14
1000.0 Ω 0 Ω -0.5 0.5 -0.7 0.7
1000.0 Ω 1 kΩ 999.0 1001.0 998.6 1001.4
10.000 kΩ 0 Ω -0.010 0.010 -0.015 0.015
10.000 kΩ 10 kΩ 9.980 10.020 9.970 10.030