753/754
Calibration Manual
22
TEKTRONIX FCA3000 Counter
Fluke 75x
Fluke 123
Use Channel A
BNC Input
gso11.eps
Figure 9. Frequency Source Verification Connections
Table 12. Frequency Source Verification Points
UUT Range Frequency @ 1 Vpp
Minimum
Fre
q
uenc
y
Maximum
Fre
q
uenc
y
10.99 Hz 5 Hz Sine 4.99 Hz 5.01 Hz
1099.9 Hz 1 kHz Sine 999.9 Hz 1000.1 Hz
21.999 kHz 10 kHz Sine 9.998 kHz 10.002 kHz
50 kHz 49 kHz Sine 48.995 kHz 49.005 kHz
10.99 Hz 5 Hz Square 4.99 Hz 5.01 Hz
1099.9 Hz 1 kHz Square 999.9 Hz 1000.1 Hz
UUT Range Frequency @ 7.5 Vpp Minimum Frequency Maximum Frequency
109.99 Hz 50 Hz Square 49.9 Hz 50.1 Hz