28
6. Continue through the test points.
7. When you complete the test, set the 5522A to STANDBY.
5522A
CALIBRATOR
Fluke 75X
Fluke 5522A
gso14.eps
Figure 12. Three-Wire RTD Measure Verification Connections
Table 16. RTD Measure Verification (100W Pt (385), Three-Wire Connection)
Input °C (Resistance) 1-Year ( °C ) 2-Year ( °C )
-180 °
(
27.096
)
-179.53 to -180.47 -179.46 to -180.54
100 °
(
138.505
)
99.53 to 100.47 99.46 to 100.54
780 ° (369.712 ) 779.39 to 780.61 779.19 to 780.81