753/754
Calibration Manual
34
of 90.0 100.000 mV 110.0., apply the requested input, or apply an input in
the allowed range. Use the numeric keys to add the value. Push the Continue
softkey.
7. Apply the subsequent requested value as in step 5. Push the Continue softkey.
When you record a negative value, always start with ‘-’.
XWWarning
Some of the voltages required for calibration are dangerous. To
avoid injury or death from electric shock, do not touch live
conductors during high-voltage calibration. Put the source
device into Standby mode after each calibration step.
8. Continue to apply voltages as necessary. Make sure you follow the connection
instructions each time. Remember that the input jack configuration changes.
9. When you complete the last point in the subroutine, you are asked if you to keep
the new constants. If you keep the new constants, the calibration constants are
saved and the Calibration Status counter is incremented and the date is updated.
If you discard the constants at this point, the calibration has no effect, the
Calibration Status counter is not incremented, and the date does not change.