Comply with local and national safety requirements
when in hazardous locations.
Only use test leads that have the same voltage,
category, and amperage ratings as the meter and
that are approved by a safety agency. When used
with optional accessories, the lowest category
rating applies.
To avoid possible electric shock, do not use
TouchHold to determine if dangerous voltage is
present. TouchHold will not capture unstable or
noisy readings.
To avoid false readings, which could lead to
possible electric shock or personal injury, replace
the battery as soon as the battery indicator (M)
appears.
To avoid electrical shock, remove test leads from
the meter before you open the battery door.
Close and latch the battery door before you use the
meter.
To avoid personal injury or damage to the meter,
use only the specified replacement fuse, 440 mA
1000 V fast-blow, Fluke PN 943121.
Do not exceed the Measurement Category (CAT)
rating of the lowest rated individual component of a
Product, probe, or accessory.
Do not use the TL175 or TP175 test probes in CAT
III or CAT IV environments without the probe tip
fully extended and correct category rating visible in
the window.
When the TL175 is used with instruments or other
accessories, the lowest category rating of the
combination applies. One exception is when the
probe is used with the AC172 or AC175.