80 Series V 
Calibration Manual 
18 
Table 14.  Accuracy Tests (cont.) 
Display Reading 
Step 
Test  
Function 
Range 5500A Output 
83 and 88  87 
23 600 Ω 330 Ω ( Use 2 wire Comp)
1
328.5 to 331.5 
329.1 to 330.9 
24 6 kΩ 3.3 kΩ (Use 2 wire Comp)
 1
3.286 to 3.314 
3.292 to 3.308 
25 60 kΩ 33 kΩ 
32.86 to 33.14 
32.92 to 33.08 
26 600 kΩ 330 kΩ 
327.6 to 332.4 
327.9 to 332.1 
27 6 MΩ 3.3 MΩ 
3.276 to 3.324 
3.279 to 3.321 
28 
e 
Ohms 
50 MΩ 30 MΩ 
29.67 to 30.33 
29.67 to 30.33 
29 
60 nS  Open input  - 0.30 to 0.30  - 0.30 to 0.30 
30 
nS 
Conductance 
60 nS  100 MΩ  9.60 to 10.40  9.60 to 10.40 
31 
G 
Diode 
6 V  3.0 V dc  2.939 to 3.061  2.939 to 3.061 
32 
? 
AC Amps 
6 A  3.0 A, 60 Hz  2.962 to 3.038  2.968 to 3.032 
33 
A 
DC Amps 
6 A  3.0 A  2.984 to 3.016  2.990 to 3.010 
33B 
A 
DC Amps
 
10A
6
  10A  9.94 to 10.06  9.96 to 10.04 
34 
60 mA  33 mA, 60 Hz  32.58 to 33.42  32.65 to 33.35 
35 
m
? 
AC Milliamps 
400 mA  330 mA, 60 Hz  325.8 to 334.2  326.5 to 333.5 
36 
60 mA  33 mA  32.83 to 33.17  32.89 to 33.11 
37 
m[  
DC Milliamp 
400 mA  330 mA  328.5 to 331.5  329.1 to 330.9 
38 
600 µA 330 µA, 60 Hz  325.8 to 334.2  326.5 to 333.5 
39 
U
?
 
AC 
Microamps 
6000 µA 3300 µA, 60 Hz  3258 to 3342  3265 to 3335 
40 
600 µA 330 µA   328.3 to 331.7  328.9 to 331.1 
41 
U[ 
DC 
Microamps 
6000 µA 3300 µA  3285 to 3315  3291 to 3309 
42 
10 nf  Open input
2
  0.21 to 0.31  0.21 to 0.31 
43 
100 nf  5 nf
5
  04.7 to 05.3  04.7 to 05.3 
44 
E 
Capacitance 
100 µf 9.5 µf  09.2 to 09.8  09.2 to 09.8 
45 
1000 V  400 V, 400 Hz  N/A  376 to 408 
46 
K 
Low Pass 
Filter 
1000 V   400 V, 800 Hz
4
  N/A  226 to 340
4