Digital Multimeter 
 
Performance Tests 
  19 
Table 14.  Accuracy Tests (cont.) 
Display Reading 
Step 
Test  
Function 
Range 5500A Output 
83 and 88  87 
47 
Max = 5.896 to 6.104 
48 
L 
(87 and 88 only) 
Peak Min/Max 
6 V dc  8 Vpp, 2 kHz Sq. 
Wave, DC offset 2 V 
Min = -1.898 to -2.102 
49 
 0 °C  -1.0 to 1.0  -1.0 to 1.0 
50 
m
L
 
(87 and 88 only) 
Temperature 
3
 
 100 °C  98.0 to 102.0  98.0 to 102.0 
51 
  Press backlight button  Backlight comes on 
52 
  Press backlight button  Backlight Intensifies 
53 
Backlight 
  Press backlight button  Backlight off 
1.   Or short test leads and use REL to offset test lead resistance. 
2.   Remove test leads from unit. 
3.   To ensure accurate measurement, the Meter and thermocouple adapter must be at the same temperature. After  
      connecting the thermocouple adapter to the Meter allow for reading to stabalize before recording display reading. 
4.   The Meter accuracy is not specified at this input signal frequency with Low-pass filter selected. The display reading  
      shown, check that the Low-pass filter is active and follows an expected roll-off curve. 
5.   Use REL to compensate for internal Meter and lead capacitance. Test leads must be disconnected from the 
calibrator before using REL. 
6.  W 10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum;  
> 10 A unspecified. 
Testing the Inductive Pickup (88 Only) 
To test the inductive pickup, a function generator output will simulate automobile spark 
plug signals on a loop of wire containing a 10 e resistor. The pickup will be clamped to 
the wire and output voltage from the pickup will be monitored by an oscilloscope.  
Perform the following procedure to test the inductive pickup: 
1.  Solder a 10-inch piece of 14 AWG wire to one end of a 10 e 1 % resistor. 
2.  Connect the other end of the 10 e resistor to the terminal LOW output of the 
function generator. Place the other end of the 14-AWG wire to the HIGH output of 
the function generator. See Figure 5. 
3.  Connect a 10X scope probe from channel 2 (dc-coupled) of the oscilloscope across 
the 10 e resistor. 
4.  Clamp the inductive pickup to the wire loop on the HIGH side of the resistor as 
shown in Figure 5. 
Make sure that the jaws of the inductive pickup are closed completely, and that the 
side of the inductive pickup that says "SPARK PLUG SIDE" points toward the 
HIGH output of the function generator. 
5.  Connect a 10X scope probe from channel 1 (dc-coupled) of the oscilloscope across 
the output of the inductive pickup.