Fluke 80i-1010
12 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1.
Set
the
voltmeter
range
to
accommodate
a
reading
of
200
mV
ac.
2.
Setthe
calibrator
and
amplifier
controls
to
source
1A
at
50
or
60
Hz
ac
inthe
50-turn
coil,
thus
producing
50A
turns.
The
voltmeter
should
display
a
reading
between
48.2
and
51.8
mV.
3.
Setthe
calibrator
and
amplifier
controls
to
source
2A
at
400
Hz
ac
in
the
50-turn
coil,
thus
producing
100A
turns.
The
voltmeter
should
display
a
reading
between
97.0
and
103.0
mV
ac.
4.
Set
the
voltmeter
range
to
accommodate
a
display
of
1.000V
ac.
5.
Setthe
calibrator
and
amplifier
controls
to
source
10A
at
50
or
60
Hz
ac
in
the
50-turn
coil,
thus
producing
500A
turns.
The
voltmeter
should
display
a
reading
between
488.0
and
512.0
mV
ac.
6.
Set
the
amplifier
to
standby.
7.
Remove
the
probe
from
the
50-turn
coil.
This
completes
the
AC
Operation
verification
procedure.
CLEANING
AND
STORAGE
Clean
the
mating
surfaces
of
the
core
at
regular
intervals.
Apply
a
light
coating
of
dripless
oil
to
prevent
corrosion.
If
the
probe
will
not
be
used
for
periods
of
longer
than
60
days,
the
batteries
should
be
removed
and
stored
separately.