TEMPERATURE RANGE
Operating O^C to 50*C
Non
Operating
-40°C to 70°C*
*0’^ C to 50^C
operating and non-operating with
calibration memory option
(
-
04)
installed.
HUMIDITY RANGE
(Operating to
Full
Accuracy)
0‘^C to 18°C 80% RH
18“C
to
40'^C
75%
RH
40‘'C to
5(fC
60%
RH
OVERLOAD
LO to
Guard 127 V max
Guard to Chassis 1000V max
HI Sense to Hi Source 127V max
LO Sense to LO Swrce
127V
max
HI Sense
to
LO Sense
1000V max
HI
Source
to
LO Source
280V max
cm
Inches
DIMENSIONS
H
L W
10.8 X
42.5 X
43.2
4.25 X
16.75
X
17
WEIGHT
Basic Fully Loaded
kg 9.1 10.92
Lbs
20 24
OPERATING POWER
Basic
instrument 12 watts 1 15V ac or 230V ac
±10%
Fully Loaded 24 watts
50-60
Hz
WARM-UP
1
hour to rated accuracy
SHOCK AND VIBRATION
Meets
requirements of
MIL-T-28800 for Class 5
style E equipment.
Note
The stated
accuracies are valid under the
following environ-
mental
conditions.
Temperature: 18'^Cto 28°C
(Except 24 hour: 22'^Cto24°C}
Humidity:
<=75%
Line Regulation: 90V to IIOV. 103.5Vto
126.5V, or 207V
to
253V @ 45 to 66 Hz
DC
Volts
Normal Resolution
(5 1/2 digits)
Full Scale
Accuracy
±
^0
of Input
+
Number of
Digits)
Range
Resulu tiuii
24 Hours 90 Days
1 Year
100 mV 312 mV 1 uV
0.002+4 0.003 +
5
0.005
+8
1
V
2.5
V 10 uV
0.001
+
1 0,002
+
1
0.004 +
1
10
V
20
V 100uV 0.001 or
1*
0.001
+
1 0.002
+
1
100 V 160
V
1 mV 0.001
+
1
0.002
+
1
0.004
+
1
1000 V
1200 V
*Whichever is greater.