Changing the Default Settings
Adjusting the Temperature Offset
5
5-5
If necessary, adjust the temperature offset until the
temperature on the primary display matches the
temperature indicated by the lag bath thermometer.
1. Press % ( s) to advance to the next digit and press
T (l) to go back to the previous digit.
2. Press [ or ] to increase or decrease the digit
value.
3. Save changes by pressing O T.
4. Press O N to exit Setup.
tc041f.eps
Figure 5-1. Adjusting Temperature Offset