Digital Multimeter
Specifications
7
Table 5. Current Function Specifications
Function Range Resolution Accuracy
1, 2
Burden Voltage (typical)
mA
\
(45 Hz to 2 kHz)
mA
[
60.00 mA
400.0 mA
4
6.000 A
10.00 A
3
60.00 mA
400.0 mA
4
6.000 A
10.00 A
3
0.01 mA
0.1 mA
0.001 A
0.01 A
0.01 mA
0.1 mA
0.001 A
0.01 A
± (1.0 % + 2)
± (1.0 % + 2)
± (1.0 % + 2)
± (1.0 % + 2)
± (0.2 % + 4)
± (0.2 % + 2)
± (0.2 % + 4)
± (0.2 % + 2)
1.8 mV/mA
1.8 mV/mA
0.03 V/A
0.03 V/A
1.8 mV/mA
1.8 mV/mA
0.03 V/A
0.03 V/A
µA B
(45 Hz to 2 kHz)
µAF
600.0 µA
6000 µA
600.0 µA
6000 µA
0.1 µA
1 µA
0.1 µA
1 µA
± (1.0 % + 2)
± (1.0 % + 2)
± (0.2 % + 4)
± (0.2 % + 2)
100 µV/µA
100 µV/µA
100 µV/µA
100 µV/µA
1. AC conversions are ac coupled, true rms responding, and valid from 3 % to 100 % of range, except 400 mA range
(5 % to 100 % of range) and 10 A range (15 % to 100 % or range).
2. The Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the
Meter may display a residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2
digit change for readings over 3 % of range. Using REL to offset this reading may produce a much larger constant
error in later measurements
3. W 10 A continuous up to 35 °C; < 20 minutes on, 5 minutes off at 35 °C to 55 °C. 20 A for 30 seconds maximum;
> 10 A unspecified.
4. 400 mA continuous; 600 mA for 18 hrs maximum.
Table 6. Capacitance and Diode Function Specifications
Function Range Resolution Accuracy
E 10.00 nF
100.0 nF
1.000 µF
10.00 µF
100.0 µF
9999 µF
0.01 nF
0. 1 nF
0.001 µF
0.01 µF
0.1 µF
1 µF
± (1 % + 2)
1
± (1 % + 2)
1
± (1 % + 2)
± (1 % + 2)
± (1 % + 2)
± (1 % + 2)
G 3.000 V 0.001 V ± (2 % + 1)
1. With a film capacitor or better, using Relative mode to zero residual.
Table 7. Frequency Counter Specifications
Function Range Resolution Accuracy
Frequency
(0.5 Hz to 200 kHz,
pulse width > 2 µs)
199.99
1999.9
19.999 kHz
199.99 kHz
> 200 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz
± (0.005 % + 1)
± (0.005 % + 1)
± (0.005 % + 1)
± (0.005 % + 1)
unspecified