Performance Tests
27
Testing the Pushbuttons
To test the pushbuttons
1. Turn the Meter rotary knob to J.
2. Press each button and note that the meter responds with a beep for each button
press.
3. Press and hold B a second time to exit MIN MAX mode.
Testing Meter Accuracy
Perform the accuracy test steps in Table 15.
Table 15. Accuracy Tests
Step Test Function Range 5500A Output Display Reading
1 600 mV 330 mV, 60 Hz 327.3 to 332.7
2 600 mV 600 mV, 13 kHz 586.0 to 614.0
3 6 V 3.3 V, 60 Hz 3.275 to 3.325
4 6 V 3.3 V, 20 kHz 3.214 to 3.386
5 60 V 33 V, 60 Hz 32.75 to 33.25
6 60 V 33 V, 20 kHz 32.14 to 33.86
7 600 V 330 V, 60 Hz 327.5 to 332.5
8 600 V 330 V, 2.5 kHz 323.0 to 337.0
9 1000 V 500 V, 60 Hz 494 to 506
10
K
AC Volts
1000 V 1000 V, 1 kHz 986 to 1014
11 600 mV 150 mV, 99.95 kHz 99.93 to 99.97
12
K
Hz
AC Volts
Frequency
600 mV 150 mV, 199.50 kHz 199.48 to 199.52
13 6 V 0.7 V, 99.95 kHz 99.93 to 99.97
14
Sensitivity
60 V 7 V, 99.95 kHz 99.93 to 99.97
15 V Hz
Trigger level
6 V 3.4 V, 1 kHz Sq. Wave 999.8 to 1000.2
16
V Hz
Duty Cycle
6 V 5 V, 1 kHz, DC offset 2.5 V
Sq. Wave
49.7 % to 50.3 %
17 6V 3.3 V dc 3.297 to 3.303
18 60 V 33 V dc 32.97 to 33.03
19 600 V 330 V dc 329.7 to 330.3
20
V
DC Volts
1000 V 1000 V dc 998 to 1002
21 600 mV 33 mV dc 32.9 to 33.1
22
m
V
DC Volts
600 mV 330 mV dc 329.6 to 330.4
23 600 330 ( Use 2 wire Comp)
1
329.1 to 330.9
24 6 k 3.3 k (Use 2 wire Comp)
1
3.292 to 3.308
25 60 k 33 k 32.92 to 33.08
26 600 k 330 k 327.9 to 332.1
27 6 M 3.3 M 3.279 to 3.321
28
e
Ohms
50 M 30 M 29.67 to 30.33