8845A/8846A
Users Manual
D-2
Table D-2. Ohms Analog Filter Errors
Range NPLC Additional Analog Filter Error
10, 100 0.5 mΩ
10 Ω
<10 1.9 mΩ
10, 100 1.5 mΩ
100 Ω
<10 9.0 mΩ
10, 100 0.6 Ω
100 kΩ
<10 2.5 Ω
Table D-3. DC Current Analog Filter Errors
Range NPLC Additional Analog Filter Error
100 μA, 10 mA, 1 A 100 0.005 % of range
10 0.015 % of range
1 0.027 % of range
0.2 0.09 % of range
0.02 0.27 % of range
1 mA, 100 mA, 10 A
[1]
10 0.001 % of range
1 0.0025 % of range
0.2 0.009 % of range
0.02 0.026 % of range
[1] 3 A range uses the 10 A range errors.