8845A/8846A 
Users Manual 
4-10 
2
INPUT SENSE
HI
LO
FRONT
 REAR
10 A
1V
00
mA
4
W
HI
LO
2W/4W
V
300V
FUSED
1000 V CAT I
  600V CAT II
4
Ground tab
 
caw060.eps 
Figure 4-3. Input Connections for 4-wire Ohms Using 2x4 Wire Leads 
 
Function modifiers: 
D FLTR  A filter for quieting noisy measurments. This filter averages readings 
to reduce reading noise when in immedicate trigger mode or when in 
trigger mode with an unending number of triggers selected. The filter 
is only available for dc functions at rates slower than 1 PLC. The 
number of readings averaged by the digital filter varies with dc 
function and range. 
A FLTR  A 3-pole analog filter for improving noise immunity. The filter is 
active when this soft key’s label is highlighted. 
Refer to the “Range Keys” section in Chapter 3 of this manual for information on how to 
adjust the measurement range.  
Measuring Current 
The Meter is capable of making both ac and dc current measurements up to 10 A. Two 
separate input connectors, in conjunction with the LO connector, are used for current 
measurements. For best resolution, current measurements not exceeding 400 mA should 
be made using the LO and mA input connectors as shown in Figure 4-4.  
W Caution 
To avoid blowing the current input fuse or possibly damaging 
the Meter:  
•  Current measurements between 400 mA and 10 A should be 
measured using only the 10 A and LO input connectors. 
•  BEFORE applying power to the circuit to be measured, 
ensure the test leads are correctly connected to the Meter 
inputs appropriate for the expected current. 
•  Exceeding 440 mA on the 400 mA input connector or 
exceeding 11 A on the 10 Amp connector will blow the 
internal fuse.