Remote Operation
Optimizing System Speed for I/O Operation
3
Note
The remote interface port and settings can be changed only through the
Meter’s front-panel controls.
Enabling or selecting a port for remote operation is a separate process from setting the
various parameters associated with a remote interface. An interface does not have to be
selected to set its parameters; for example, the baud rate of the RS-232 port can be set
while the Meter has the LAN port enabled as the remote control port.
Optimizing System Speed for I/O Operation
Several factors affect the remote system speed of the 8845A/46A. To make the fastest
measurements, it is best to store the measurements in internal memory before transferring
across the I/O bus. This is particularly true if you want to collect a large number of
samples per measurement. Set up all of the parameters (i.e. function, range, filters, etc...)
prior to taking the measurements. Use the internal trigger. Set the trigger delay to zero.
Turn off the display. If operating with RS-232, use the fastest baud rate available.
One example of taking fast voltage readings is:
*cls # This clears all error prior to initiating readings
conf:volt:dc 0.1 # Set the DC range to manual and 100 mV.
volt:dc:nplc 0.02 # Set the NPLC to faster reading rate 4 1/2 digits.
zero:auto 0 # Turn Autozero off.
trig:sour imm # Set immediate trigger.
trig:del 0 # Set trigger delay to zero.
trig:coun 1 # Set trigger count to one or specify a larger number.
disp off # Turn display off. This may slightly improve reading rate.
syst:rem # Put the 8846A in remote. Only needed for RS-232 or Ethernet
#(required for READ? or MEAS?)
samp:coun 100 # Set the sample count from 1 to 5000.
:INIT; *OPC? # "*OPC?" returns 1 indicating that the measurements have been
taken and are ready to be transferred to the computer using the
FETCH? command.
:FETCH?
Note
For any sample count greater than 1, measurements are returned comma
separated.

Table of Contents

Other manuals for Fluke 8846A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 8846A and is the answer not in the manual?

Fluke 8846A Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range10 Ω to 1 GΩ
ContinuityYes
Diode TestYes
Temperature MeasurementYes
Display5.5 digit
ConnectivityUSB, RS-232, GPIB
DC Current Range100 µA to 10 A
AC Current Range100 µA to 10 A
Frequency Accuracy0.01 %
Capacitance Range1 nF to 100 µF
Display TypeVFD
Power Supply100 V to 240 V
AC Voltage Accuracy±(0.06% + 0.03% of reading)
Resistance Accuracy±(0.01% + 0.001% of reading)

Related product manuals