3. Press “SET” to program in the new set-point.
4. Press and hold “EXIT” to return to the temperature display.
When the set-point temperature is changed the controller switches the well heater on or
off to raise or lower the temperature. The displayed well temperature gradually changes
until it reaches the set-point temperature. The well may require 5 to 10 minutes to reach
the set-point depending on the span. Another 5 to 10 minutes is required to stabilize
within ±0.1°C of the set-point. Ultimate stability may take 15 to 20 minutes more of
stabilization time.