Fluke 9100S - Set-Point Value; Temperature Scale Units; Scan

Fluke 9100S
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1 100. Set-pointmemory1location,100°Ccurrentlyused
To change to another set-point memory press the up or down arrow.
4 300. Newset-pointmemory4location,300°C
Press “SET” to accept the new selection and access the set-point value. Press “EXIT” to
continue and to ignore any changes.
S
Acceptselectedset-pointmemory
6.2.2 Set-point Value
The set-point value may be adjusted after selecting the set-point memory and pressing
“SET”.
4 200. Set-point4valuein°C
If the set-point value does not need to be changed, press and hold “EXIT” to resume
displaying the well temperature. To change the set-point value, press “SET” and then
press the up or down arrow.
220.0 Newset-pointvalue
When the desired set-point value is reached, press “SET” to accept the new value and
access the temperature scale units selection. If “EXIT” is pressed, any changes made to
the set-point are ignored.
S
Acceptnewset-pointvalue
6.2.3 Temperature Scale Units
The temperature scale units of the controller maybe set by the user to degrees Celsius
(°C) or Fahrenheit (°F). The units are used in displaying the well temperature, set-point,
proportional band, and high limit.
Press “SET” after adjusting the set-point value to change display units.
Un= C Scaleunitscurrentlyselected
Press the up or down arrow to change the units.
Un= F Newunitsselected
Press “SET” to accept the new units or “EXIT” to cancel.
Note: The temperature scale units may also be changed by pressing “SET” and u when
the temperature is displayed. This action toggles the units between °F and °C.
6.3 Scan
The scan rate can be set and enabled so that when the set-point is changed the dry-well
heats or cools at a specied rate (degrees per minute) until it reaches the new set-point.
With the scan disabled the dry-well heats or cools at the maximum possible rate.