BTT=−
32
C
TTTT
=
4433
100
1
100 100
1
100
D
TTTT
=
3322
100
1
100 100
1
100
ER R=−
43
FRR=−
32
delta
AF BE
DE CF
=
T
1-4
- Measured temperature using thermometer.
R
1-4
- Value of R from display of 9103 (Press SET and DOWN at the same
time.)
where
T
1
and R
1
are the measured temperature and resistance at –25°C
T
2
and R
2
are the measured temperature and resistance at 0°C
T
3
and R
3
are the measured temperature and resistance at 60°C
T
4
and R
4
are the measured temperature and resistance at 125°C
10.2.2 Compute R0 and ALPHA
a T delta
TT
12
22
100
1
100
=+
a T delta
TT
34
44
100
1
100
=+
rzero
Ra Ra
aa
=
41 23
13
alpha
RR
Ra Ra
=
24
41 23
delta is the new value of DELTA computed above.