Wireless AC Clamp 
  Performance Tests 
7 
Backlight 
To verify that the backlight functions: 
1.  With the Product ON, push . 
2.  The backlight will come on. If it does not, repair is necessary. See 
Contact Fluke. 
Keypad Test 
To verify that the keypad functions, turn ON the Product and push each button 
separately. Each button push will turn on a display annunciator, and  will turn 
on the backlight. If the buttons do nothing, repair is necessary. See 
Contact Fluke. 
AC Current Test 
Before you do the ac current test: 
1.  Make sure that you have the necessary equipment. See Table 2.  
2.  Make sure the Product battery is good and replace it if necessary. See 
Battery Replacement. 
3.  Warm up the Calibrator as necessary. Refer to its specifications. 
4.  Let the temperature of the UUT become stable to room temperature. 
To do the ac current test: 
1.  Connect the Calibrator A ac output and ground to the 50-Turn Coil. See 
Figure 3. 
2.  Apply the input level for each step shown in Table 3. 
3.  Compare the indication on the Product display with the UUT reading limits in 
Table 3. 
4.  If the display indication falls outside of the range shown in Table 3, calibration 
adjustment or repair of the Product is necessary. See Calibration Adjustment. 
Table 3. Performance Tests 
Test 
Calibrator 
Output 
Resolution Specification 
UUT Reading Limit 
Low High 
AC Amps 
(with 50-
Turn Coil) 
0.2 A,50 Hz 
0.1 
2.0 % 
9.3 10.7 
1 A,50 Hz  48.5  51.5 
2 A,50 Hz  97.5  102.5 
4 A,50 Hz  195.5  204.5 
8 A,50 Hz  391.5  408.5 
0.2 A,150 Hz 
2.5 % 
9.25 10.75 
4 A,150 Hz  194.5  205.5 
8 A,150 Hz  389.5  410.5 
0.2 A,400 Hz  9.25  10.75 
2 A,400 Hz  97  103 
4 A,400 Hz  194.5  205.5 
8 A,400 Hz 
389.5 410.5