Wireless AC Clamp 
  Before Calibration Adjustment 
11 
Restore the Default Password 
If the calibration password is lost, the default password (1234) can be manually 
restored with the subsequent steps: 
 Warning 
To prevent electric shock or personal injury, remove all input 
signals before you open the Product. 
1.  Remove the Product battery door. See Battery Replacement.   
2.  With a Phillips screwdriver, remove the bottom case screws. Two of the 
screws are inside of the battery door. 
3.  Keep the pca in the top case. 
4.  Apply 3.0 V across the battery contacts on the pca. Note the polarity that is 
shown in Figure 4. 
5. Push  on the front of the Product. 
6.  Short across the CAL keypad on the pca. See Figure 4. The default password 
is now restored. 
7.  Remove the 3.0 V supply and replace the bottom case, batteries, and battery 
door.   
Short across calibration
key pads to reset default 
password
Battery 
contacts
 
hby08.eps 
Figure 4. Calibration Password Reset