EasyManua.ls Logo

Fluke CableAnalyzer DSP-4000SR - Scanning Function; When to Use a Remote Unit

Fluke CableAnalyzer DSP-4000SR
228 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DSP-4000 Series
Users Manual
4-2
Scanning Function
The Single Test versions of the wire map, resistance, HDTDR, and HDTDX
analyzer tests include a scanning function, which you can activate by pressing the
# Scanning ON softkey. The scanning function runs the test repeatedly and
updates the display each time a test is complete. This function is useful for finding
intermittent problems on a cable.
Note
To extend battery life, connect the ac adapter/charger when using
the scanning function for more than 1 minute.
When to Use a Remote Unit
A remote unit is required only when testing twisted pair cable. Table 4-1 shows
which cable tests require a remote and which remotes support each test.
If a remote unit is detected at the start of a Single Test, the test tool runs a wire
map test before running the selected test. If the wire map test fails, the test tool
stops the test and displays the wire map. Press $ Continue Test to run
the selected test.
Note
The DSP-4100 test tool is compatible only with DSP-4100SR units.
The DSP-4100SR unit is not compatible with the DSP-100,
DSP-2000, or DSP-4000 test tools.