EasyManua.ls Logo

Fluke CombiScope PM3380B - Page 207

Fluke CombiScope PM3380B
297 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMMAND REFERENCE 4 - 107
SYSTem:TIME
Syntax:
SYSTem:TIME <hour>,<minute>,<second>
<hour> <NRf> | MINimum | MAXimum
Range from 0 to 23.
<minute> <NRf> | MINimum | MAXimum
Range from 0 to 59.
<second> <NRf> | MINimum | MAXimum
Range from 0 to 59.
Query form:
SYSTem:TIME? [MINimum | MAXimum , MINimum | MAXimum ,
MINimum | MAXimum]
Response:
<hour>,<minute>,<second>
The time values returned are of type <NR1>. If MINimum was
specified, the lowest possible value is returned. If MAXimum was
specified, the highest possible value is returned.
Description:
The SYSTem:TIME command programs the real-time clock of the instrument by
specifying the hour, minute, and second. Only a 24-hours time format is
supported. The current time is not changed after a
*
RST command.
Example:
Send
SYSTem:TIME 11,22,33 Sets the system time to 11
hours + 22 minutes + 33
seconds, i.e.: 11:22:33.
Send
SYSTem:TIME? MAX,MAX,MAX Queries for the max. values
possible.
Read
23,59,59 Reads 23 hours, 59 minutes,
59 seconds.
Front panel compliance:
The SYSTem:TIME command is the remote equivalent of the UTILITY - CLOCK
- hh:mm:ss softkey menu.

Table of Contents