DTX-MFM2/GFM2/SFM2 Fiber Modules
Users Manual
22
-continued-
Autotest in Smart Remote Mode: Set the Reference
1
Turn on the tester and smart remote and let them sit
for 5 minutes. Allow additional time if the modules
have been stored above or below ambient
temperature.
2
Turn the rotary switch to SETUP, then select Fiber Loss.
Set the following on the Fiber Loss tabs (press
C to
see other tabs):
Test Limit: Select the test limit required for the job.
Press
J More to see other lists of limits.
Fiber Type: Select the fiber type to be tested.
Remote End Setup: Set to Smart Remote.
Bi-directional: Enable this if you are required to test
the fiber in both directions.
Number of Adapters and Number of Splices: Enter
the number of adapters and splices that will be
added to each direction of the fiber path after the
reference is set.
Connector Type: Select the connector type used in the
cabling to be tested. Select General if the exact type is
not listed.
Test Method: Refers to the number of adapters
represented in the loss results. Select 1 Jumper if you
use the reference and test connections shown in this
manual.
3
Turn the rotary switch to SPECIAL FUNCTIONS, then
select Set Reference. If both a fiber module and a
twisted pair or coaxial adapter are attached, select Fiber
Module next.
4
The Set Reference screen shows reference connections
for the test method you selected. Figure 8 shows
connections for the 1 Jumper method. Clean and
inspect the connectors on the testers and test reference
cords, connect the tester and smart remote; then press
P.
See Table 2 on page 15 for details on settings.