SPECIFICATIONS
Table
1-1.
Specifications (cont)
FUNCTION
RANGE
RESOLUTION
VAC
3.200V
0.001V
(45-500
Hz, 3.2V rng
32.00V
0.01V
Other
ranges
45-1
kHz)
320.0V
0.1V
750V
IV
o
320.00
0.10
3200 o
1 o
32.00 kO 0.01
kO
320.0 kO
0.1 kO
3.200 MO 0.001
MO
32.00
MO
0.01 MO
-H-
'«!)
2.0V 0.001V
AAC*
32.00
mA** 0.01 mA
(45
Hz
to 1 kHz)
320
mA** 0.1 mA
10.00
A*** 0.01 A
ADC*
32.00 mA**
0.01 mA
320 mA**
0.1 mA
10.00
A*** 0.01A
*
Current function
not
available on Fluke 70
**
Ranges available
on Fluke 75/21 and 77/23 only.
***
10A
continuous,
20A for
30 seconds
maximum
BURDEN
VOLTAGE
(TYPICAL)
Not
Applicable
FUNCTION
RESOLUTION
ACCURACY
70 73/75/21
77/23
0.001V
±(0.5%+1
)
+(0.4%+1)
±(0.3%+1)
0.01V
±(0.5%+1)
±(0.4%+1
)
±(0.3%+1)
0.1V
±(0.5%+1)
±(0.4%+1)
±(0.3%+1
)
IV
±(0.6%+1)
±(0.4%+1
)
±(0.4%+1)
0.1 mV
±(0.5%+1) ±(0.4%+1
)
±(0.3%+1
)

Other manuals for Fluke II Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke II Series and is the answer not in the manual?

Fluke II Series Specifications

General IconGeneral
True RMSYes
Temperature-40 °C to 400 °C
ContinuityYes
Diode TestYes
Safety RatingCAT III 600V
Voltage DC600V
Voltage AC600V
Current DC10A
Current AC10A
Resistance40MΩ
Battery Life400 hours
Warranty3 years