Fluke 19xB-19xC-2x5C
Service Manual
Possible error messages.
The following messages can be shown on the test tool display:
WARNING: Calibration data not valid.
Save data and exit maintenance mode?
Proceed as follows:
If you did the WarmingUp and Pre-Calibration successfully (section 5.5), and you
want to store the Pre-Calibration data before continuing with the Final Calibration:
Press
YES
.
When turning the test tool off and on again, it will show the message:
The instrument needs calibration.
Please contact your service center.
The calibration date and number will not be updated. You must continue with the Final
Calibration!
To return to the Maintenance mode, if you want to repeat the complete calibration:
Press
NO
.
Now press until the display shows
WarmingUp (CL 0200):IDLE
, and
calibrate the test tool, starting at section 5.5.
If you want to exit and maintain the old calibration data:
Turn the test tool off.